Please use this identifier to cite or link to this item: https://www.um.edu.mt/library/oar/handle/123456789/101678
Title: Optimal design of electrodes for functional electrical stimulation applications to single layer isotropic tissues
Authors: Cassar, Mary Grace
Sebu, Cristiana
Pidcock, Michael
Chandak, Shubham
Andrews, Brian
Keywords: Electrodes -- Design and construction -- Mathematical models
Electrodes -- Surfaces -- Design and construction
Electric stimulation -- Mathematical models
Fredholm equations -- Numerical solutions
Issue Date: 2023
Publisher: Emerald
Citation: Cassar, M. G., Sebu, C., Pidcock, M., Chandak, S., & Andrews, B. (2023). Optimal design of electrodes for functional electrical stimulation applications to single layer isotropic tissues. COMPEL- The International Journal for computation and mathematics in electrical and electronic engineering, 42(3), 695-707.
Abstract: Purpose – The purpose of this paper is to investigate the design of skin surface electrodes for functional electrical stimulation using an isotropic single layered model of the skin and underlying tissue. A concentric ring electrode geometry was analysed and compared with a conventional configuration, specifically to localise and maximise the activation at depth and minimise the peak current density at the skin surface. Design/methodology/approach – The mathematical formulation determines the spatial electric potential distribution in the tissue, using the solution to the Laplace equation in the lower half space subject to boundary conditions given by the complete electrode model and appropriate asymptotic decay. Hence, it is shown that the electric potential satisfies a weakly singular Fredholm integral equation of the second kind which is then solved numerically in MATLAB for a novel concentric ring electrode configuration and the conventional two disk side-by-side electrode configuration. Findings – In both models, the electrode geometry can be optimised to obtain a higher activation and lower maximum current density. The concentric ring electrode configuration, however, provides improved performance over the traditional two disk side-by-side electrode configuration. Research limitations/implications – In this study, only a single layer of medium was investigated. A comparison with multilayer tissue models and in vivo validation of numerical simulations are required. Originality/value – The developed mathematical approaches and simulations revealed the parameters that influence nerve activation and facilitated the theoretical comparison of the two electrode configurations. The concentric ring configuration potentially may have significant clinical advantages.
URI: https://www.um.edu.mt/library/oar/handle/123456789/101678
Appears in Collections:Scholarly Works - FacSciMat



Items in OAR@UM are protected by copyright, with all rights reserved, unless otherwise indicated.