Title: Maltese CT doses for commonly performed examinations demonstrate alignment with published DRLs across Europe
Authors: Zarb, Francis
McEntee, Mark
Rainford, Louise
Keywords: Abdomen -- Anatomy
Radiation -- Dosage -- Standards
Radiation -- Health aspects
Radiography, Medical -- Image quality
Chest -- Radiography
X-rays -- Safety measures
X-rays -- Equipment and supplies
Radiography, Medical -- Equipment and supplies
Issue Date: 2012
Publisher: Oxford University Press
Citation: Zarb, F., McEntee, M., & Rainford, L. (2012). Maltese CT doses for commonly performed examinations demonstrate alignment with published DRLs across Europe. Radiation protection dosimetry, 150(2), 198-206.
Abstract: This work recommends dose reference levels (DRLs) for abdomen, chest and head computerised tomography (CT) examinations in Malta as the first step towards national CT dose optimisation. Third quartiles volume CT dose index values for abdomen: 12.1 mGy, chest: 13.1 mGy and head: 41 mGy and third quartile dose-length product values for abdomen: 539.4, chest: 492 and head: 736 mGy cm–1 are recommended as Maltese DRLs derived from this first Maltese CT dose survey. These values compare well with DRLs of other European countries indicating that CT scanning in Malta is consistent with standards of good practice. Further work to minimise dose without affecting image quality and extending the establishment of DRLs for other CT examinations is recommended.
URI: https://www.um.edu.mt/library/oar/handle/123456789/106827
Appears in Collections:Scholarly Works - FacHScRad

Files in This Item:
File Description SizeFormat 
Maltese_CT_doses_for_commonly_performed_examinations_demonstrate_alignment_with_published_DRLs_across_Europe.pdf
  Restricted Access
168.33 kBAdobe PDFView/Open Request a copy


Items in OAR@UM are protected by copyright, with all rights reserved, unless otherwise indicated.